Tiers 1-3 (Dealer Manual) 900 SeriesCAUTION: USA Federal law restricts this device to sale by or on the order of a physician.22090 Rev. CRemRest™
EC REPMDSS GmbHSchigraben 4130175 Hannover, Germany
SPECIFICATIONSSIZE4.7" H X 8.6" W X 9.4" D11.9cm H x 21.8cm W x23.9cm DPOWER110-240VAC, 50-60Hz,12V DC inputPRESSURE RANGE3.5 cmH2O To
RemRest Set UpTier 1:To enter the Technician Mode, plug one end of the AC power cordinto the AC inlet plug located on the back of the RemRest device.P
TO SET THE PRESCRIPTION PRESSURE(BLINK ONCE/OFF 3 SECONDS):The default setting is 10cmH2O. Using a manometer and pushingthe up and down arrows, adjust
NOTE: Tier 1 model does not incorporate automatic altitude compensation. In order to insure that this device provides theexpected pressures, the follo
Tier 2&3:Note: There are no differences between the Tier 2 and Tier 3 labelother than color. Tier 2’s label is aqua and Tier 3’s label is blu
RemRest 7
TO VIEW OR RESET THE USAGE TIME:T